Comparison of refractive outcomes using Scheimpflug Holladay equivalent keratometry or IOLMaster 700 keratometry for IOL power calculation

dc.contributor.authorAksoy, Mustafa
dc.contributor.authorAsena, Leyla
dc.contributor.authorGungor, Sirel Gur
dc.contributor.authorKucukoduk, Ali
dc.contributor.authorAkman, Ahmet
dc.contributor.orcID0000-0001-8024-4758en_US
dc.contributor.pubmedID33733281en_US
dc.date.accessioned2022-09-14T10:12:52Z
dc.date.available2022-09-14T10:12:52Z
dc.date.issued2021
dc.description.abstractPurpose This study aims to compare postoperative refractive error results using Pentacam (Oculus Optikgerate GmbH) Holladay equivalent keratometry readings (EKR) or IOLMaster 700 (Carl Zeiss Meditec AG) keratometry (K) values in IOL power calculation. Material and methods This retrospective study included 54 eyes of 31 patients who underwent cataract surgery. Preoperative biometric measurements of all patients were obtained using IOLMaster 700 followed by Pentacam measurements. IOLMaster 700 K measurements on horizontal (K1) and vertical (K2) axes and EKR measurements on 2 mm (EKR2mm), 3 mm (EKR3mm) and 4.5 mm (EKR4.5 mm) corneal zones were recorded. EKR4.5 mm value and IOLMaster 700 K values were used in Holladay-II, SRK/T, Haigis, and Hoffer-Q formulas to calculate predictive refractive error (PRE). Absolute refractive error (ARE) was calculated as the absolute difference between actual postoperative refractive error (APRE) and PRE values. Results Mean age was 72.2 +/- 8.3 (51-87) years and mean IOL power was 21.5 +/- 2.9 D (18-23 D). There was no significant difference between PRE values when IOLMaster 700 K measurements and EKR4.5 mm K measurements were used in Holladay-II, SRK/T, Haigis, and Hoffer-Q formulas (p = 0.571, p = 0.833, p = 0.165, p = 0.347, respectively). There was no significant difference between APRE and ARE values (p = 0.124). According to mean ARE results, the closest estimate was achieved when the IOLMaster 700 K values were used in the Holladay-II formula (p = 0.271). Conclusion IOLMaster 700 K measurement and Pentacam EKR4.5 mm measurements can be used interchangeably. IOLMaster 700 K values yielded the most predictive measurement of the refractive result using the Holladay-II formula.en_US
dc.identifier.endpage2212en_US
dc.identifier.issn0165-5701en_US
dc.identifier.issue6en_US
dc.identifier.scopus2-s2.0-85102555775en_US
dc.identifier.startpage2205en_US
dc.identifier.urihttp://hdl.handle.net/11727/7730
dc.identifier.volume41en_US
dc.identifier.wos000629929800001en_US
dc.language.isoengen_US
dc.relation.isversionof10.1007/s10792-021-01781-6en_US
dc.relation.journalINTERNATIONAL OPHTHALMOLOGYen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergien_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectOptical Biometryen_US
dc.subjectIOLMaster 700en_US
dc.subjectPentacamen_US
dc.subjectCataract surgeryen_US
dc.subjectEquivalent keratometry readingsen_US
dc.titleComparison of refractive outcomes using Scheimpflug Holladay equivalent keratometry or IOLMaster 700 keratometry for IOL power calculationen_US
dc.typearticleen_US