User Verification of Abbott Alinity HQ Hematology Analyzer

dc.contributor.authorBayraktar, Nilufer
dc.contributor.authorTopcu, Deniz Ilhan
dc.contributor.orcID0000-0002-1219-6368en_US
dc.contributor.orcID0000-0002-7886-3688en_US
dc.contributor.researcherIDE-3717-2019en_US
dc.contributor.researcherIDY-8758-2018en_US
dc.date.accessioned2024-05-10T11:40:22Z
dc.date.available2024-05-10T11:40:22Z
dc.date.issued2023
dc.description.abstractObjectives This study aims to evaluate the performance characteristics of the Alinity HQ hematology analyzer in a routine laboratory setting.Methods In the study, precision (short-term and long-term precision), accuracy (method comparison with Abbott Cell Dyn Ruby and estimation of bias), confirmation of a background (Limit of Blank, LoB), and carry-over were used to evaluate the performance of Alinity HQ as recommended by ICSH, CLSI guidelines EP15-A3, EP09, EP17A2, and H26-A2. Acceptance criteria were based on manufacturer technical specifications and the EFLM Biological Variation Database.Results According to the short-term precision results, except for mean corpuscular volume (MCV) and mean corpuscular hemoglobin concentration (MCHC), all measurements exhibited coefficient variations (CV) lower than their verification limits. Basophil, eosinophil, and monocyte counts, as well as mean corpuscular hemoglobin (MCH), MCHC, and red cell distribution width standard deviation (RDW-SD), did not meet the allowable imprecision criteria for the long-term precision study. The estimated bias for all analytes was within verification limits. However, the method comparison study showed concentration-dependent variations for MCHC, MCH, and mean platelet volume (MPV) parameters. Furthermore, the correlation of parameters between Alinity HQ and Cell Dyn Ruby ranged from 0.46 to 1.00. The LoB and carry-over studies demonstrated satisfactory performance for the Alinity HQ analyzer.Conclusions Although some parameters had higher CVs than expected and concentration-dependent bias, the overall analytical performance of Alinity HQ was found to be satisfactory. Alinity HQ is an accurate, highly precise analyzer with good analytical performance, suitable for high-volume laboratories.en_US
dc.identifier.eissn1303-829Xen_US
dc.identifier.endpage177en_US
dc.identifier.issn0250-4685en_US
dc.identifier.issue2en_US
dc.identifier.startpage167en_US
dc.identifier.urihttps://www.degruyter.com/document/doi/10.1515/tjb-2023-0185/html
dc.identifier.urihttp://hdl.handle.net/11727/12086
dc.identifier.volume49en_US
dc.identifier.wos001118830500001en_US
dc.language.isoengen_US
dc.relation.isversionof10.1515/tjb-2023-0185en_US
dc.relation.journalTURKISH JOURNAL OF BIOCHEMISTRY-TURK BIYOKIMYA DERGISIen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergien_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectmethod verificationen_US
dc.subjecthematology analyzeren_US
dc.subjectCLSI H26-A2en_US
dc.subjectICSH guidelineen_US
dc.subjectCLSI EP15-A3en_US
dc.titleUser Verification of Abbott Alinity HQ Hematology Analyzeren_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
10.1515_tjb-2023-0185.pdf
Size:
2.03 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: