Restoration of pull-out strength of the failed pedicle screw: biomechanical comparison of calcium sulfate vs polymethylmethacrylate augmentation

dc.contributor.authorGuler, Umit Ozgur
dc.contributor.authorDerincek, Alihan
dc.contributor.authorHersekli, Murat Ali
dc.contributor.authorOzalay, Metin
dc.contributor.authorCinar, Bekir Murat
dc.contributor.authorAcaroglu, Emre
dc.date.accessioned2019-12-23T13:14:34Z
dc.date.available2019-12-23T13:14:34Z
dc.date.issued2014
dc.description.abstractObjective: The aim of the present study was to compare calcium sulfate (CAS) and polymethylinethaaylate (PMMA) bone cements used for the augmentation of a failed pedicle screw with biomechanical pull-out strength (POS) testing. Methods: Thirty lumbar vertebrae were harvested from 6 calves and bone mineral densities (BMD) were measured. Primary polyaxial pedicle screws were randomly inserted and pulled out and the POSs of the specimen were recorded. For revision, specimens were randomly assigned to the CAS-augmented pedide screws group (Group 1) or PMMA-augmented pedicle screw group (Group 2). Pull-out tests were repeated to compare both groups. Results: Mean BMD of the specimens was 1.006 +/- 0.116 g/cm(2). There were no statistically significant differences between BMD results of the two groups (p=0.116). For Group 1, mean POS of primary screws was 2,441.3 +/- 936.4 N and was 2,499.5 +/- 1,425.1 N after CAS augmentation, demonstrating no statistically significant difference (p=0.865). In Group 2, mean POS of the primary screws was 2,876.6 +/- 926.6 N and significantly increased to 3,745.5 +/- 1,299.2 N after PMMA augmentation (p=0.047). There was also a significant difference in mean POS between the CAS and PMMA groups (p=0.026). Conclusion: Although CAS augmentation facilitates a revision screw POS as strong as that of primary screws, it is not as strong as PMMA augmentation.en_US
dc.identifier.endpage206en_US
dc.identifier.issn1017-995X
dc.identifier.issue2en_US
dc.identifier.startpage202en_US
dc.identifier.urihttps://pdfs.semanticscholar.org/7258/431fe4e030e5ab458eb2739499cfe211f6f5.pdf?_ga=2.57695850.1597073326.1577106751-1432498723.1577106751
dc.identifier.urihttp://hdl.handle.net/11727/4526
dc.identifier.volume48en_US
dc.identifier.wos000335199800016
dc.language.isoengen_US
dc.relation.isversionof10.3944/AOTT.2014.3193en_US
dc.relation.journalACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICAen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectCalcium sulfateen_US
dc.subjectpedide screwen_US
dc.subjectPMMAen_US
dc.subjectpull-out strengthen_US
dc.titleRestoration of pull-out strength of the failed pedicle screw: biomechanical comparison of calcium sulfate vs polymethylmethacrylate augmentationen_US
dc.typearticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Thumbnail Image
Name:
431fe4e030e5ab458eb2739499cfe211f6f5.pdf
Size:
1.18 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: