Comparison of Intraocular Pressure Measurements Between Goldmann Applanation Tonometry and Reichert 7 Noncontact Tonometry

dc.contributor.authorGungor, Sirel G.
dc.contributor.authorAkman, Ahmet
dc.contributor.authorYazici, Ayse C.
dc.contributor.orcID0000-0002-3132-242Xen_US
dc.contributor.orcID0000-0001-6178-8362en_US
dc.contributor.pubmedID24145288en_US
dc.contributor.researcherIDAAS-6810-2021en_US
dc.contributor.researcherIDAAD-5967-2021en_US
dc.date.accessioned2024-01-16T12:51:31Z
dc.date.available2024-01-16T12:51:31Z
dc.date.issued2015
dc.description.abstractPurpose:The aim of this study was to compare the Reichert 7 (R7) noncontact tonometry with Goldmann applanation tonometry (GAT) and to determine the influence of central corneal thickness (CCT) in these measurements in the healthy population.Design:A prospective cross-sectional study.Methods:Intraocular pressure (IOP) of the right eyes of 120 patients was measured with GAT and R7. All patients were free of glaucoma. All the measurements were carried out between 7 am and 9 am The measurements with the R7 were taken in the automatic mode. After 15 minutes, IOP with GAT was measured followed by pachymetry. The R7 provided a Goldmann-correlated IOP (IOPg) and a corneal-compensated IOP (IOPcc).Results:The mean age of the patients was 53.919.3 (range, 26 to 85 y). The male/female ratio was 0.9/1. When R7 measurements were compared with GAT measurements, R7 measurements (both IOPcc and IOPg) were significantly higher than GAT measurements (P0.001). Besides that, IOPcc was significantly higher than IOPg (P0.001). There was a significant positive linear relationship between IOPcc and GAT (r=0.761, P0.001), and similar relationship between IOPg and GAT (r(2)=0.739, P0.001). The mean CCT was 545.9 +/- 33.2 m (range, 476 to 634 m). A weak correlation was observed between CCT and GAT (r=0.196, P=0.032). There was a significant correlation between CCT and IOPg (r=0.283, P=0.02). The correlation between CCT and IOPcc was not statistically significant (r=0.123, P=0.179).Conclusions:IOP values of R7 are higher than GAT values. However, IOPg was observed more coherent to the GAT values than IOPcc. IOPcc should be an evaluation factor along with GAT or IOPg in glaucoma examination.en_US
dc.identifier.eissn1536-481Xen_US
dc.identifier.endpage441en_US
dc.identifier.issn1057-0829en_US
dc.identifier.issue6en_US
dc.identifier.scopus2-s2.0-84939573038en_US
dc.identifier.startpage438en_US
dc.identifier.urihttp://hdl.handle.net/11727/11269
dc.identifier.volume24en_US
dc.identifier.wos000360296700008en_US
dc.language.isoengen_US
dc.relation.isversionof10.1097/IJG.0000000000000009en_US
dc.relation.journalJOURNAL OF GLAUCOMAen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergien_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectReichert 7 noncontact tonometryen_US
dc.subjectGoldmann applanation tonometryen_US
dc.subjectcentral corneal thicknessen_US
dc.titleComparison of Intraocular Pressure Measurements Between Goldmann Applanation Tonometry and Reichert 7 Noncontact Tonometryen_US
dc.typeArticleen_US