Evaluation of the possible effect of magnetic resonance imaging noise on peripheral hearing organ with the otoacoustic emission
No Thumbnail Available
Date
2020
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Purpose: The aim of this study is to evaluate the effect of noise produced by magnetic resonance imaging (MRI) device on hearing by using objective and subjective audiological assessments.
Methods: A total of 38 patients between the ages of 18 and 50 without hearing loss, and had performed MRI for brain, head, neck or cervical imaging were included in this prospective clinical study. Pure tone audiometry, speech audiometry, high frequency audiometry, transient evoked otoacoustic emissions (TEOAE) and distortion product otoacoustic emission (DPOAE) were performed before and after MRI.
Results: There was no statistically significant difference in TEOAE, pure tone audiogram, high frequency audiogram and speech audiogram thresholds. In DPOAE, the median value before and after MRI at the frequency of the left ear at 4.0 kHz was 13.6 (8.5-19.9) and 15.7 (8.9-20.7) SNR respectively (p > .05). The median value before MRI at the right ear 4.0 kHz frequency was 14.1 (9.1-20.5) SNR, whereas the median value after MRI was 13.2 (8.8-19.8 SNR (p = 0,03). There was no statistically significant difference in other frequencies in DPOAE.
Conclusions: This is the first objective study that examines the MRI noise on speech audiometry and otoacoustic emission together. However, the effect of MRI noise on hearing pathway is still doubt. Based on the difference at 4 kHz frequency on DPOAE; on-earphones may not sufficiently protect the patients from the MRI noise and this issue should deserve further research.
Description
Keywords
Magnetic resonance imaging (MRI), Hearing, High frequency audiometry, Transient Evoked Otoacoustic Emissions (TEOAE), Distortion Product Otoacoustic Emissions (DPOAE)