Non-Contact and Contact Tonometry in Corneal Edema
No Thumbnail Available
Date
2016
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Purpose To compare the intraocular pressure (IOP) values by Goldmann applanation tonometry (GAT) and Reichert 7 CR noncontact tonometry (R7CR-NCT) in patients with postsurgical corneal edema and to examine the impact of postoperative corneal edema on these values.
Methods Forty-six patients with grade 4 and 5 cataracts were included in this study. Intraocular pressure was measured using GAT and R7CR-NCT before and 1 day after phacoemulsification. Central corneal thickness (CCT) was determined before and after surgery to quantify postsurgical corneal edema. The R7CR-NCT provided a Goldmann-correlated IOP (IOPg) and corneal-compensated IOP (IOPcc).
Results The CCT increased significantly 1 day after surgery (the mean preoperative CCT, 543.5 25.4 m; the mean postoperative CCT, 681.5 +/- 19.8 m; p < 0.001), a mean increase of 26.7%. The preoperative R7CR-NCT measurements (the mean IOPcc, 18.8 +/- 5.6 mm Hg; the mean IOPg, 17.8 +/- 4.5 mm Hg) were significantly higher than GAT measurements (the mean GAT-IOP, 16.0 +/- 3.4 mm Hg) (p < 0.001). On postoperative day 1, the R7CR-NCT measurement (the mean IOPcc, 27.0 +/- 9.8 mm Hg; the mean IOPg, 25.1 +/- 8.9 mm Hg) were significantly higher than GAT measurements (the mean GAT-IOP, 18.3 +/- 7.9 mm Hg) (p < 0.001). The difference between postoperative R7CR-NCT and GAT-IOP values were significantly higher than the difference between preoperative R7CR-NCT and GAT-IOP values (p < 0.001 for both IOPcc to GAT-IOP and IOPg to GAT-IOP). The difference between postoperative IOPcc and GAT-IOP was significantly correlated to the change in CCT (r = 0.526, p < 0.001). Similarly, the difference between postoperative IOPg and GAT-IOP was significantly correlated to the change in CCT (r = 0.536, p < 0.001).
Conclusions The R7CR-NCT IOP values were high in patients with postsurgical intensive corneal edema. Consequently, the difference between R7CR-NCT IOP and GAT-IOP increased in edematous corneas.
Description
Keywords
Goldmann applanation tonometry, Reichert 7 CR noncontact tonometry, Goldmann-correlated intraocular pressure, corneal-compensated intraocular pressure, corneal edema