Feasibility of Carbon-Fiber-Reinforced Polymer Fixation Plates for Treatment of Atrophic Mandibular Fracture: A Finite Element Method

dc.contributor.authorDiker, Nurettin
dc.contributor.authorBurak, Bayram
dc.contributor.orcID0000-0002-7825-1083en_US
dc.contributor.pubmedID30340836en_US
dc.contributor.researcherIDAAS-4519-2020en_US
dc.date.accessioned2023-05-09T08:14:53Z
dc.date.available2023-05-09T08:14:53Z
dc.date.issued2018
dc.description.abstractThe objective of this study was to conduct a computer assessment of the biomechanical stability of locking fixation plates of different thicknesses, made of titanium alloy and carbon-fiber- reinforced polyetheretherkotone (CFR-PEEK) in Class III atrophic mandibular fractures. Class III atrophic mandibular models were constructed using three-dimensional finite element models. After simulation of fracture on the left side, plates with different thicknesses (1.0, 1.5, 2.0, and 2.5 mm) were adapted to the models and three locking screws on each side of the fracture were used for fixation. Titanium alloy and CFR-PEEK material properties were assessed for all plate models. Von Mises stress values decreased gradually with an increase in plate profile thickness. Von Mises stress values for screws and plates were lower in models using CFR-PEEK plates. An increase in plate profile enhanced the load-sharing performance of the fixation systems, with rigid titanium alloy plates sharing the load among the screws more evenly. 1.0 mm, 1.5 mm, and 2.0 mm CFR-PEEK plates produced strain values of 2-10%, which promote proper healing via formation of callus in the fracture line. In terms of investigated biomechanical parameters, and with a Young's modulus similar to that of cortical bone, CFR-PEEK materials appear to be suitable for the treatment of atrophic mandibular fractures. (C) 2018 European Association for Cranio-Maxillo-Facial Surgery. Published by Elsevier Ltd. All rights reserved.en_US
dc.identifier.endpage2189en_US
dc.identifier.issn1010-5182en_US
dc.identifier.issue12en_US
dc.identifier.scopus2-s2.0-85054865610en_US
dc.identifier.startpage2182en_US
dc.identifier.urihttp://hdl.handle.net/11727/8931
dc.identifier.volume46en_US
dc.identifier.wos000452705400027en_US
dc.language.isoengen_US
dc.relation.isversionof10.1016/j.jcms.2018.09.030en_US
dc.relation.journalJOURNAL OF CRANIO-MAXILLOFACIAL SURGERYen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergien_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectAtrophic mandibular fractureen_US
dc.subjectInternal fixationen_US
dc.subjectCarbon fiberen_US
dc.subjectFixation plateen_US
dc.subjectFinite element methoden_US
dc.titleFeasibility of Carbon-Fiber-Reinforced Polymer Fixation Plates for Treatment of Atrophic Mandibular Fracture: A Finite Element Methoden_US
dc.typearticleen_US

Files

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: