Evaluation of maxillary trabecular microstructure as an indicator of implant stability by using 2 cone beam computed tomography systems and micro-computed tomography
dc.contributor.author | Kulah, Kivanc | |
dc.contributor.author | Gulsahi, Ayse | |
dc.contributor.author | Kamburoglu, Kivanc | |
dc.contributor.author | Geneci, Ferhat | |
dc.contributor.author | Ocak, Mert | |
dc.contributor.author | Celik, H. Hamdi | |
dc.contributor.author | Ozen, Tuncer | |
dc.contributor.pubmedID | 30595501 | en_US |
dc.date.accessioned | 2021-02-28T15:32:21Z | |
dc.date.available | 2021-02-28T15:32:21Z | |
dc.date.issued | 2019 | |
dc.description.abstract | Objective. The aim of this study was to assess the trabecular microarchitecture of the maxilla by using cone beam computed tomography (CBCT) and micro-computed tomography (micro-CT) ex vivo. Study Design. Seventeen maxillary cadaver specimens were scanned by using micro-CT and CBCT devices. Samples were scanned with 2 CBCT devices at different voxel sizes (0.08, 0.125, and 0.160 mm for 3-D Accuitomo 170; 0.75 and 0.200 mm for Planmeca Promax 3-D Max). Morphometric parameters, such as bone volume/total volume (BV/TV) ratio, trabecular thickness (Tb.Th), trabecular separation (Tb.Sp), trabecular number (Tb.N), and degree of anisotropy (DA) were assessed by using CTAnalyzer software. Bland-Altman limits of agreement and intraclass correlation coefficient (ICC) were performed to evaluate agreement between CBCT and micro-CT in consideration of measured morphometric parameters. Statistical significance was set at P < .05. Results. The BV/TV, Tb.Th, Tb.Sp, and DA values were higher for CBCT images compared with micro-CT images, whereas the Tb. N value was lower with CBCT images than with micro-CT images. The BV/TV and DA parameters showed the highest agreement between CBCT and micro-CT devices (ICC = 0.421 for BV/TV and ICC = 0.439 for DA; P < .01). Conclusions. The BV/TV and DA parameters measured on CBCT obtained at the smallest voxel size were found to be useful for the assessment of maxillary trabecular microstructure. | en_US |
dc.identifier.endpage | 256 | en_US |
dc.identifier.issn | 2212-4403 | en_US |
dc.identifier.issue | 3 | en_US |
dc.identifier.scopus | 2-s2.0-85059145708 | en_US |
dc.identifier.startpage | 247 | en_US |
dc.identifier.uri | http://hdl.handle.net/11727/5447 | |
dc.identifier.volume | 127 | en_US |
dc.identifier.wos | 000458263100012 | en_US |
dc.language.iso | eng | en_US |
dc.relation.isversionof | 10.1016/j.oooo.2018.11.014 | en_US |
dc.relation.journal | ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | BONE-MINERAL DENSITY | en_US |
dc.subject | MORPHOMETRIC-ANALYSIS | en_US |
dc.subject | ACCURACY | en_US |
dc.subject | PARAMETERS | en_US |
dc.subject | CT | en_US |
dc.title | Evaluation of maxillary trabecular microstructure as an indicator of implant stability by using 2 cone beam computed tomography systems and micro-computed tomography | en_US |
dc.type | article | en_US |
Files
License bundle
1 - 1 of 1
No Thumbnail Available
- Name:
- license.txt
- Size:
- 1.71 KB
- Format:
- Item-specific license agreed upon to submission
- Description: