Comparison of 2 Devices in Pigs To Induce Hypothermia in Laparoscopic Orthotopic Kidney Transplant

dc.contributor.authorHan, Xiuwu
dc.contributor.authorZhang, Yuhai
dc.contributor.authorGao, Qiang
dc.contributor.authorZhao, Zhiwei
dc.contributor.authorYan, Wei
dc.contributor.authorZhang, Bao
dc.date.accessioned2026-04-14T06:50:31Z
dc.date.issued2012-12
dc.description.abstractObjectives: To laparoscopically compare the effectiveness of 2 cooling devices for renal hypothermia and investigate the feasibility of laparoscopic orthotopic kidney transplant using a pig model. Materials and Methods: Eight pigs were divided into 2 groups of 4 animals each. Laparoscopic nephrectomy and autotransplant were performed first on only the right kidney. One week later, these procedures were performed on the left kidney, while the first transplanted autograft was removed. After 1 more week, the left autograft was removed for observation. In 1 group, the silicon tube cage was used to induce hypothermia during laparoscopic orthotopic kidney transplant (silicon tube cage group), and in the other group, the plastic bag jacket was used to induce hypothermia during laparoscopic orthotopic kidney transplant (plastic bag jacket group). Results: Two pigs in the silicon tube cage group survived for 7 days after the second autotransplant with serum creatinine levels of 210 µmol/L and 1010 µmol/L. One pig in the plastic bag jacket group survived for 5 days. The mean surface temperature of the grafts was maintained at 9°C ± 3°C and 12°C ± 3°C in the silicon tube cage and plastic bag jacket groups (P = .166). Three of 6 plastic bag jacket devices were ruptured by stitches or instruments. The mean venous and arterial anastomotic times in the silicon tube cage group were significantly shorter than were those in the plastic bag jacket group. Conclusions: Our study shows that the silicon tube cage may be a reliable renal cooling device for use in laparoscopic kidney transplant and indicates the feasibility of laparoscopic orthotopic kidney transplant in pigs.
dc.identifier.citationExperimental and Clinical Transplantation, Cilt, 10, Sayı, 6, 2012 ss. 573-578en
dc.identifier.eissn2146-8427en
dc.identifier.issn1304-0855
dc.identifier.issue6en
dc.identifier.urihttps://hdl.handle.net/11727/14930
dc.identifier.volume10en
dc.language.isoen
dc.publisherBaşkent Üniversitesi
dc.sourceExperimental and Clinical Transplantationen
dc.subjectKidney
dc.subjectHeterotopic
dc.subjectOrthotopic transplant
dc.subjectLaparoscopy
dc.subjectHypothermia devices
dc.titleComparison of 2 Devices in Pigs To Induce Hypothermia in Laparoscopic Orthotopic Kidney Transplant
dc.typeArticle

Files

Original bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
573.pdf
Size:
404.8 KB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: